MapXpert和CheckMate3西林样品残氧(顶空气体)检测对比
发布时间:2024-06-25来源:上海锦川机电技术有限公司
MapXpert和CheckMate3西林样品检测对比
检测视频:
https://v.douyin.com/i6jrwSsC/ 05/24 T@L.wf QXm:/
上海锦川的MapXpert采用CheckMate 3完全兼容的氧化锆氧传感器,性能和CheckMate 3 完全一致。MapXpert氧化锆氧传感器可直接用在CheckMate 3.
Zirconia O2 sensor:
Measure time Min. 3 sec. (incl. 1 sec. measure delay time)1
Sample volume Min. 1,35 ml (with a 3 sec measuring time and a flow of
27 ml/min)
Typical sample volume 2,83 ml ( with a 5 sec measuring time and a flow of
34 ml/min)
Reference gas Ambient atmospheric air (20.9 % O2)
1 This time should be increased when the sample pressure is lower than atmospheric pressure.
Accuracy specification (excl. calibration):
Zirconia sensor resolution 0,1 % absolute in range above 10 %
0,01 % absolute in range above 1 %
0,001 % absolute in range below 1 %
Zirconia sensor accuracy ±0,01 % absolute in range below 1 %
±1 % relative in range above 1 %
为此,我们做了西林瓶实际样品检测。
检测方式:
手动检测模式
检测结果:
平均值 | ||||||
CheckMate 3 | 2.63 | 2.70 | 2.49 | 2.56 | 2.47 | 2.57 |
MapXpert | 2.47 | 2.53 | 2.77 | 2.74 | 2.70 | 2.642 |
MapXpert Data Sheet
-