TR047 残氧仪(电化学氧传感器)的响应时间
发布时间:2023-09-25来源:上海锦川机电技术有限公司
残氧仪(电化学氧传感器)的响应时间
PDF版本下载:
https://cloud.189.cn/web/share?code=uiiiuabQvUf2
(访问码:cpf2)
本试验主要通过MapScan检测纯氮,验证和说明使用电化学氧传感器的残氧仪(顶空气体分析仪)的响应时间。同时,我们对有些厂家虚标参数,进行说明。同时,对实际检测的建议。
MapScan使用电化学氧传感器,此传感器和几乎所有进口仪器使用同一供应商的氧传感器。
氧传感器参数如下:
- Electrochemical
- Range: 0-100%
- Resolution: 0.1%
- Accuracy:
‘0.25% abs. + 2% rel. @ CO2 < 25%
‘0.9% abs. + 2% rel. @ CO2 >= 25%
- Temperature compensation:
< 2% rel. (0-40 oC)
Response time (T95):<= 10 sec.
Expected life-time: > 1 year (approx. 2 years in 20.9% O2)
Measuring system:
- Sample flow 90-140 ml/min standard
- Measuring time:
10 sec. (Fixed, CheckPoint O2)
6-10 sec. (Adjustable, MapScan O2)
- Built-in pump max. -200 mbar vacuum, expected life-time > 500 hours (180,000 measurements)
- Calibration: Air (20.9% O2)
检测视频说明
MapScan响应时间视频:
https://cloud.189.cn/web/share?code=bqIZZrBvyQNv
(访问码:6mok)
检测对象 | 检测结果 | 说明 | |
1 | 空气 | 20.8% | |
2 | 100% N2 | 0.4% | |
3 | 100% N2 | 0.1% | |
4 | 100% N2 | 0.0% | |
5 | 空气 | 20.6% | |
6 | 空气 | 20.7% |
说明:
仪器的检测时间为10秒,仪器的流量为120ml/min。我们发现检测完空气,直接检测氮气,需要检测3次后才会出现正确的读值0.0%。这个主要由于电化学氧传感器的特性决定,这个不是由仪器决定。任何进口的仪器都会出现这个现象。氧传感器需要足够的样品气体量和足够的时间响应。
在本检测可以发现,需要检测3次才能读到准确的值。如果氧含量1.0%,可能需要1次或2次检测后,就可以读取正确的值。
最小样品气体量
某些厂家标出手持电化学氧残氧仪最小样品气体量为2ml。这个数值完全接近氧化锆氧传感器。众所周知,氧化锆氧传感器性能远超电化学氧传感器。通过,这个试验可以观察2ml样品气体是远无法达到正常的残氧检测。
实际产品检测的建议:
案例A:
比方产品实际的氧含量1.0%左右,出厂的氧含量标准为小于1.8%。实际产品实际氧含量和品控标准有较大空间。如果要求不是太高,第一次检测就可作为实际检测结果。
案例B:
脱氧剂厂家需要看实际的脱氧剂的效果。0.3%和0.0%是2个完全不同的结果。对于类似这种产品,建议多次检测,并做好记录。